Ambient Air Quality Monitoring

NOx Analyzer – Z-105i Ein

Model Z-105i Ein

Z-105i Ein chemiluminescence NOx analyzer measuring NO, NO₂ and NOx from ppb to ppm for ambient air monitoring and dilution-method CEMS.

  • Designed & manufactured in-house
  • Tested & calibrated to traceable national standards
  • Installation, AMC & after-sales support

Talk to our technical team: 08048057962

Overview

The Z-105i Ein measures nitrogen oxides using chemiluminescence detection, with a modular design and microprocessor control. It reports NO, NO₂ and total NOx independently, from low ppb levels up to ppm.

Selectable measuring ranges, an ozonator with a permeable drying tank, timed dark-current compensation and adaptive signal filtering support stable readings and fast response. Large internal memory stores historical data automatically, and bidirectional RS232 allows remote control.

It is designed for atmospheric (ambient) monitoring and can also be used in continuous emission monitoring systems that use the dilution method.

Key features

  • Chemiluminescence detection
  • Independent NO, NO₂ and NOx readings
  • ppb to ppm measuring capability; selectable ranges
  • Ozonator with permeable drying tank
  • Timed dark-current compensation and adaptive filtering
  • Automatic historical data storage
  • RS232 / RS485 communication with remote control

Applications

  • Ambient air quality monitoring stations
  • Dilution-method CEMS
  • Environmental research and compliance monitoring

Technical specifications

Specifications
Measuring range 0~500ppb, 0~20,000ppb (adjustable)
Unit ppb, ppm, ug/m3, mg/m3
Zero noise <0.2ppb(RMS)
Limit of detection 0.4ppb
Zero drift <1ppb/24h
Span drift <5ppb/24h
Response time <120s (T90)
Linearity <1%/F.S.
Repeatability <1%
Sampling flow range 400SCCM±50SCCM
Ambient temperature 5℃~40℃
Dimension 178mmx430mmx635mm
Weight <16kg, external pump of 5kg
Power 200V~240V, 50/60Hz, 400W
Analog output 10V, 5V, 1V, 0.1V (selectable)
Communication RS232, RS485
I need